Biosafety Cabinets and HVAC Design

Fume hoods, pharmacy barrier isolators and some types of biosafety cabinets require a remote exhaust system to remove potentially hazardous particulates and/or gases and vapors from the lab environment.

The following information and links may be useful for those who need to design an HVAC system prior to purchasing and installing new biosafety cabinets.

Construction of Biosafety Cabinet Exhaust Systems

Of course, ductwork construction must be strong enough to withstand the static pressure of the lab equipment in order to prevent distortion or collapse. Additionally, personnel protection depends on a steady flow of air, so it’s important to limit the number of turns or bends between the biosafety cabinet exhaust connection and the roof exhaust blower. Class II Type B1 and B2 cabinets require a bubble tight damper (also known as an air tight damper) at or near the equipment connection. This damper seals off the cabinet from the building exhaust ductwork during gas decontamination and is also used by certifiers when testing the exhaust airflow alarm.

Dedicated Exhaust Systems

NSF/ANSI Standard 49 Annex E recommends that all Class II Type B1 and B2 cabinets have a dedicated exhaust system to minimize potential fluctuations in airflow and pressure that may result in a loss of containment at the cabinet work access opening. However, many constant air volume valves are being used in manifolded design facilities with varying results. These automated valves have been shown to maintain proper exhaust airflow during static pressure fluctuations that would typically cause a loss in cabinet face velocity.

Measuring Biosafety Cabinet Exhaust Airflow

Biosafety manufacturers should provide two exhaust requirements for their biosafety cabinets, fume hoods and pharmacy barrier isolators: exhaust airflow volume and duct static pressure. Maintaining these values is critical for proper cabinet operation, especially for Class II Type B2 cabinets, which require more exhaust airflow and static pressure compared to other ventilation equipment. Each value should be verified prior to equipment installation (unfortunately, this is often neglected).

The exhaust static pressure measurement should be taken near the equipment/duct connection, typically no more than 2 duct diameters away – not at the building exhaust fan. A duct traverse is the traditional method for obtaining the exhaust airflow rate. Per ASHRAE 111-2008 Standards, readings should be taken a minimum of 7.5 duct diameters downstream of any turns or bends and 2.5 duct diameters upstream of any turns or bends, respectively. These measurements are then converted to obtain the exhaust airflow rate.

For Class II Type B1 and B2 biosafety cabinets, manufacturers are required to provide a concurrent balance value, or exhaust flow measurement, after the biosafety cabinet has been airflow certified using a Direct Instrument Measurement (DIM) device and a thermal anemometer.

What Is a Concurrent Balance Value?

The Concurrent Balance Value was implemented by NSF Joint Committee members to minimize exhaust airflow discrepancies occurring in the field. This value is determined using the duct traverse measurement method as outlined above. Prior to determining the concurrent balance value, it must be confirmed that the cabinet is operating at its nominal set points for inflow and downflow velocity within 3 FPM. Inflow velocity is set using a DIM, the accuracy of which must be at least +/- 3% and +/- 7 CFM. The static pressure is also measured approximately two duct diameters from the cabinet exhaust connection, and filter load and tolerance values are added to the base static pressure to account for filter loading (0.3″ W.G. for Type B1 biosafety cabinets and 0.7″ W.G. for Type B2 cabinets).

The resulting values may be used for design and balance of exhaust/supply HVAC requirements.

Additional Resources

In addition to NSF/ANSI Standard 49 Annex E and ASHRAE Standard 111-2008, the following resource may be useful:

You May Also Be Interested In:

BSC Technical Training | Baker Company | Oct 19-21, 2026

Read more about BSC Technical Training | <span translate='no'>Baker</span> Company | Oct 19-21, 2026

Baker Flow Werkbanken – Kwaliteit die blijft presteren

Read more about <span translate='no'>Baker</span> Flow Werkbanken – Kwaliteit die blijft presteren

Baker Hosts International Dealer Meeting & Factory Visit in Bridgend, Wales

Read more about <span translate='no'>Baker</span> Hosts International Dealer Meeting &amp; Factory Visit in Bridgend, Wales

First EuroFlow EF-N Installed in the U.S. at University of Miami Sylvester Comprehensive Cancer Center

Read more about First EuroFlow EF-N Installed in the U.S. at University of Miami Sylvester Comprehensive Cancer Center

Cold Lab Services and The Baker Company: A Strong New Partnership

Read more about Cold Lab Services and The <span translate='no'>Baker</span> Company: A Strong New Partnership

Lifecycle Support Update 2026: Baker and/or CleanAir Benelux Flowbenches

Read more about Lifecycle Support Update 2026: <span translate='no'>Baker</span> and/or CleanAir Benelux Flowbenches

The Baker Company certified ISO 14001:2015 for 2026

Read more about The <span translate='no'>Baker</span> Company certified ISO 14001:2015 for 2026

How Hypoxia Rewires Epigenetics and Splicing to Drive EMT in Breast Cancer

Read more about How Hypoxia Rewires Epigenetics and Splicing to Drive EMT in Breast Cancer

What Oxygen Really Means for Extracellular Vesicles

Read more about What Oxygen Really Means for Extracellular Vesicles

Stay up to date